Testing of a high-contrast imaging coronagraph based on an active liquid crystal array | |
Xi Zhang; Deqing Ren; Yongtian Zhu; Jiangpei Dou; Xue Wang | |
2011-12-05 | |
会议录名称 | Optical Instruments and Technology: Optical Systems and Modern Optoelectronic Instruments |
卷号 | 8197 |
期号 | 8197 |
页码 | Vol.8197 0M-1-10 |
会议日期 | 2011-11-6 |
会议地点 | Beijing, China |
会议主办者 | The International Society for Optical Engineering |
出版者 | SPIE |
摘要 |
Liquid crystal modulator is an active optical component that is promising to replace traditional passive optical components. For high-contrast imaging coronagraphs that are used for direct imaging of extra-solar planets, passive coronagraph optical components are often adopted. It is impossible to actively optimize such a coronagraph system to achieve its best performance. Thus we’ve proposed a novel high-contrast imaging coronagraph which uses a liquid crystal array (LCA) for pupil apodizing. In our test, the LCA is well calibrated for amplitude errors and amplitude nonuniformity with the entire coronagraph optics. Close-loop compensations are applied according to the amplitude
calibration results. By doing so, a contrast of 10-4or 10-5can be achieved in an angular distance down to 3~5λ/D, which can be used for the direct imaging for young and Jupiter-like planets. The contrast can be further improved if a deformable mirror (DM) is deployed to correct wave-front errors induced by the LCA and the coronagraph optics. |
关键词 | Exoplanets Detection High-contrast Imaging Liquid Crystal Array Amplitude Correction Non-uniformity Compensation |
学科领域 | 系外行星探测 |
文献类型 | 会议论文 |
条目标识符 | http://ir.niaot.ac.cn/handle/114a32/626 |
专题 | 会议论文 |
推荐引用方式 GB/T 7714 | Xi Zhang,Deqing Ren,Yongtian Zhu,et al. Testing of a high-contrast imaging coronagraph based on an active liquid crystal array[C]:SPIE,2011:Vol.8197 0M-1-10. |
条目包含的文件 | 下载所有文件 | |||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
Testing of a high-co(863KB) | 开放获取 | CC BY-NC-ND | 浏览 下载 |